欢迎来到 登录 未经易推广诚信审核

Shantou Keyi instrument & Equipment

产品介绍

Piezo balance particle/dust monitor Mode

价 格:¥电议

型 号:

产品完善度:

生产地:其他访问量:534次

发布日期:2009/11/2 0:00:00

更新日期:1900/1/1 0:00:00

详细内容

 

 

applications

★ accuracy is ensured to meet the latest needs of dust measuring in extremely clean indoor environments

★cleanroom monitoring pharmaceutical, aerospace/defense electronics, mems, semiconductor food processing medical/ hospitals.

★cleanrooms iaq investigations food industry filter testing aerospace hospital surgical rooms paint spray booths

 

principle

 

as air sample enters the system, it travels through the impactor, which captures and removes larger particulates away from the sample.  smaller particulates become electrically charged and deposited on the piezo-crystal. 

 the total mass of the deposited particulates affects the piezo-crystal’s frequency.  since the change in frequency is proportional to the mass of the particulates, the actual weight of the particulates is obtained

 

features

  • direct measurement of aerosol mass by kanomax unique piezobalance method
  • precise. no need to compensate readings regardless of particle character
  • zeroing not needed
  • se-lectable run duration (24s or 120s)
  • built-in cleaning mechanism for easy maintenance of piezobalance transducer
  • calculation of aerosol mass concentration from cpm
  • powered by built-in ni-cd battery or ac adapter
  • while conventional dust meters "count" particulates, the piezobalance dust meter "weighs" mass concentration of particulates.
  • data can be reviewed on-screen, printed or downloaded to a computer.
  • software included for easy data download and processing.
  • easy operation requires no special training

specifications:

measuring ob-ject

airborn particles < φ10 µm

measuring range

mass concentration of 0.02-10mg/m3

sampling flow rate

1liter/min

measuring time

se-lected by switch.  built-in timer

high concentration level (1 to 10 mg/m3): 24 sec

low concentration level (0 to 1 mg/m3): 2 minutes

accuracy

+/-10% of reading +/-0.01mg/m3

cleaning frequency

cleaning is required after every 10-20 measurements

warning indicator on the display. the impactor is to be cleaned daily

power supply

built-in rechargeable ni-cd battery or ac adapter

dimensions

12.2” x 6.7” x 5.1” (311 x 170 x 130mm)

weight

approx. 8.8lbs. (2kg)

 

提交询价