欢迎来到 登录 未经易推广诚信审核

Shantou Keyi instrument & Equipment

产品介绍

Digestion / Extraction System

价 格:¥电议

型 号:XT-9900 model automatic microwave

产品完善度:

生产地:其他访问量:676次

发布日期:2009/11/2 0:00:00

更新日期:1900/1/1 0:00:00

详细内容

XT-9900 model automatic microwave microwave Digestion / Extraction System
streamlined dissolving sample apparatus for chemistry lab

 

1、main specifications and performance:
electricity requirement:220V±10% 50Hz A.C
power:1360W
microwave power:800W
microwave frequency:2450Hz
pressure setting range:0.1Mpa-4Mpa increment:0.1Mpa
Adjustable microwave power :50-100P increment: 10%P
timing adjust range:1-9999s increment:1s
max working temperature:250℃

2、SC60-1 high pressure dissolving tin.

Max control pressure: 4 Mpa s
max working temperature: 200℃
outside of the tin has PPS inter underlay

 capacity :60mL(specifications:60-240 mL )
fit for mid pressure and low pressure dissolving.

Features

 

●double CPU,automatic controlling pressure and adapting to  working pressue.
●high anti-jamming ability, immune from anti-jamming of microwave and visible light, can check and run at any time.
●having the function of pausing, infrared remote control and limiting pressure raising rate.
●can set 9 operation program at one time, auto saving.
●working procedure can divide into 9 steps, each step can auto work and auto switch.
●menu operating , simply and easy to learn.
●displaying clearly.
●automatic complete calibrate zero and full.

 

safety precautions
having functions of pressure auto control , limiting pressure raising and pressure display.
Over-pressure automatic alarming, and the gas from the tin will exclude off.
The shield of operating panel can make the user safety.
Urgent pushbutton for switch off electricity instantly.
Remote controller can stop the microwave processor.

 

XT-III model microwave processor with automatic pressure sealing

 —essential dissolving sample apparatus for analytical lab

Product quality according with the enterprise standard of shanghai city Q/IMDE1-2002
apparatus features:

Multiple patented mechanism for safety protection (over-pressure release, high temperature alarm, etc.)

The shield of operating panel can make the user safety.
auto alarming when the Digestion / Extraction time over.

Specifications
electricity requirement
A.C 220V±10% 50Hz
power
1360W
microwave power
800W
microwave frequency
2450MHz
time setting range
1-35min
max operating pressure is 4Mpa.
there are 8 pressure setting between 0- 4Pmpa. The pressure and temperature can be set according to the sample. The max working temperature of container is 200

the high pressure can also be fit for low and middle pressure.

提交询价